34 1290 Infinity Quaternary Pump User Manual
3 Installing the Module
Optimizing the Stack Configuration
Optimizing the Stack Configuration
If your module is part of a complete Agilent 1290 Infinity Liquid
Chromatograph, you can ensure optimum performance by installing the
following configurations. These configurations optimize the system flow
path, ensuring minimum delay volume.
For other possible configurations, please refer to the Agilent 1290 Infinity
System Manual.
One Stack Configuration
Ensure optimum performance by installing the modules of the Agilent
1290 Infinity Quaternary LC System in the following configuration (see
Figure 8 on page 35 and Figure 9 on page 36). This configuration
optimizes the flow path for minimum delay volume and minimizes the
bench space required.
The Agilent 1290 Infinity Quaternary Pump should always be installed at
the bottom of the stack.