134 1290 Infinity Quaternary Pump User Manual
9 Maintenance
Replacing the Inlet Weaver
Replacing the Inlet Weaver
Parts required p/n Description
G4204-81090 1290 Infinity Quaternary Pump Inlet Weaver Assembly
Preparations Switch off pump at the main power switch
Remove the front cover
Use an optional solvent shutoff valve or lift up solvent filters inside solvent bottles for avoiding
leakages
For easy access to the inlet weaver assembly, remove tubing connections between MCGV and
degasser
1 Open the screw at the bottom of the inlet valve. 2 Open the fitting at the center of the multi-channel
gradient valve (MCGV). Do not open the screw marked
with the red cross. Remove the inlet weaver from the
MCGV.
Fitting