1290 Infinity Quaternary Pump User Manual 167
Maintenance
9
Disassembling the Secondary Pump Head
10
To dry the pump head and to remove residues, use
pressurized air ((1), (2) and (3)).
11 If the wash seal shall be replaced by a new one, use the
steel side of the insert tool for removing it.
&
(
'
Wash seal