170 1290 Infinity Quaternary Pump User Manual
9 Maintenance
Replacing the Heat Exchanger
5
If removed, first insert the spacer fitting. Then insert the
new heat exchanger to the opening in the pump head and
lift it over the pins.
6 Use the 19 mm screw for fixing the front plate.
7 Insert and fix the screw.
Pins
Spacer fitting
Front plate
Screw