1290 Infinity Quaternary Pump User Manual 187
Maintenance
9
Replacing the Multi Purpose Valve
The central (C) port is connected to the outlet of the pressure sensor.
Port 1 is connected to the outlet of the optional Jet Weaver
Port 2 is connected to the inlet of the optional Jet Weaver
Port 3 is blocked by a blank nut (plastic)
Port 4 is connected to the system (typically autosampler)
Port 5 is connected to the outlet of the inline filter
Port 6 is blocked by a blank nut (SST)
Port 7 is connected to the waste capillary
Port 8 is connected to the inlet of the inline filter
Block unused ports with blank nuts.