EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1290

Agilent Technologies 1290 User Manual

Agilent Technologies 1290
302 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #243 background imageLoading...
Page #243 background image
1290 Infinity Quaternary Pump User Manual 243
Hardware Information
12
Firmware Description
Firmware Updates
Firmware updates can be done using the following tools (latest version
should be used):
Agilent Lab Advisor software with files on the hard disk
(*)
Firmware Update Tool with local files on the hard disk
(*)
Instant Pilot (G4208A) with files on a USB Flash Disk
(*)
Required tools, firmware and documentation are available from the
Agilent web:
http://www.chem.agilent.com/_layouts/agilent/downloadFirmware.aspx?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
PPPP is the product number, for example, 1315B for the G1315B DAD,
R the firmware revision, for example, A for G1315B or B for the G1315C
DAD,
VVV is the revision number, for example 650 is revision 6.50,
XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware
in chapter "Maintenance" or use the documentation provided with the
Firmware Update Tools.
Figure 32 Firmware Update Mechanism
NOTE
Update of main system can be done in the resident system only. Update of the resident
system can be done in the main system only.
Main and resident firmware must be from the same set.
GZh^YZciHnhiZb
BV^c;LjeYViZ
BV^cHnhiZb
GZh^YZci;LJeYViZ

Table of Contents

Other manuals for Agilent Technologies 1290

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 and is the answer not in the manual?

Agilent Technologies 1290 Specifications

General IconGeneral
BrandAgilent Technologies
Model1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals