26 1290 Infinity Quaternary Pump User Manual
2 Site Requirements and Specifications
Site Requirements
Bench Space
The module dimensions and weight (see Table 1 on page 27) allow you to
place the module on almost any desk or laboratory bench. It needs an
additional 2.5 cm (1.0 inches) of space on either side and approximately
8 cm (3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the
bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Condensation
CAUTION
Condensation within the module
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.