266 1290 Infinity Quaternary Pump User Manual
13 LAN Configuration
Initialization Mode Selection
Since the default IP address is a so- called local address, it will not be
routed by any network device. Thus, the PC and the module must reside
in the same subnet.
The user may open a Telnet session using the default IP address and
change the parameters stored in the non- volatile memory of the module.
He may then close the session, select the initialization mode Using Stored,
power- on again and establish the TCP/IP connection using the new
parameters.
When the module is wired to the PC directly (e.g. using a cross- over cable
or a local hub), separated from the local area network, the user may
simply keep the default parameters to establish the TCP/IP connection.
NOTE
In the Using Default mode, the parameters stored in the memory of the module are not
cleared automatically. If not changed by the user, they are still available, when switching
back to the mode Using Stored.

Table of Contents

Other manuals for Agilent Technologies 1290

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 and is the answer not in the manual?

Agilent Technologies 1290 Specifications

General IconGeneral
BrandAgilent Technologies
Model1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals