1290 Infinity Quaternary Pump User Manual 29
Site Requirements and Specifications
2
Performance Specifications
Composition precision <0.15 % RSD, or 0.02 min SD, whatever is
greater (1 mL/min).
Based on retention
time at constant room
temperature
Composition accuracy ±0.40 % absolute (1 – 99 % B, 0.5 –
2.0 mL/min with water/caffeine tracer,
400 bar)
Integrated degassing unit Number of channels: 4
Internal volume per channel: 1.5 mL
Control Agilent ChemStation for LC (C.01.04 or
above)
OpenLAB (A.04.04)
Masshunter (B.05.01 or above)
Local control Agilent Instant Pilot (G4208A) (B.02.08 or
above)
Communications Controller-area network (CAN), RS232C,
APG remote: ready, start, stop and
shutdown signals, LAN
Safety and maintenance Extensive diagnostics, error detection and
display through Agilent LabAdvisor, leak
detection, safe leak handling, leak output
signal for shutdown of the pumping
system. Low voltage in major
maintenance areas.
GLP features Early maintenance feedback (EMF) for
continuous tracking of instrument usage
in terms of seal wear and volume of
pumped mobile phase with pre-defined
and user settable limits and feedback
messages. Electronic records of
maintenance and errors.
Housing All materials recyclable.
Tabl e 2 Performance specifications
Type Specification Comments