EasyManua.ls Logo

Agilent Technologies 16900A

Agilent Technologies 16900A
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20 16900A, 16902A, and 16903A Service Guide
3 Testing Performance
Testing Performance
To ensure the logic analysis system is operating correctly,
you can perform software tests (self-tests) on the system.
Directions for performing the software tests are given in this
chapter. Because there are no specifications for the
16900- series logic analysis system, there are no performance
verification tests.
Self-Tests
The self- tests listed in this chapter check the functional
operation of the mainframe. Self- tests for the optional
modules installed in the frames are listed in the individual
module Service Guides.
There are two types of self-tests: self-tests that automatically
run at powerup, and self- tests that you select on the screen.
For descriptions of the tests, refer to Chapter 8, “Theory of
Operation.”
Perform the self- tests as an acceptance test when receiving
the logic analysis system or when the logic analysis system
is repaired.
If a test fails, refer to Chapter 5, “Troubleshooting.”
Test Interval
There is no recommended test interval for the 16900- series
mainframes. However, each of the supported modules has
performance verification tests and therefore require a
periodic verification of specifications. Refer to the Service
Guides of the individual modules for more information.