EasyManua.ls Logo

Agilent Technologies 16900A

Agilent Technologies 16900A
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Testing Performance 3
16900A, 16902A, and 16903A Service Guide 21
To perform the power-up tests
The logic analysis system automatically performs power-up
tests when you apply power to the instrument. Any errors
are reported in the boot dialogue. Serious errors will
interrupt the boot process.
The power- up tests are designed to complement the
instrument on- line Self- Tests. Tests that are performed
during powerup are not repeated in the Self-Tests.
A monitor (16900A only), keyboard and mouse must be
connected to the mainframe to observe the results of the
power- up tests.
1 Exit all logic analysis sessions.
2 Press the power switch for about 1 second and release,
wait for the the unit to power off.
3 After a few seconds, turn the power switch back on.
Observe the boot dialogue for the following:
Ensure all of the installed memory is recognized
Any error messages
Interrupt of the boot process with or without error
message
A complete transcript of the boot dialogue is in chapter 8,
"Theory of Operation."
4 During initialization, check for any failures.
If an error or an interrupt occurs, refer to Chapter 5,
“Troubleshooting.”

Table of Contents

Related product manuals