114 34410A/11A/L4411A Service Guide
4 Disassembly and Repair
7 Remove front panel.
a Using a small bladed screwdriver, gently pry the black terminal latch
from the red terminal housing. Rotate the Terminal larch up and
remove it from the instrument.
b Remove the Torx screw holding the front panel assembly.