36 34410A/11A/L4411A Service Guide
1 Specifications
Configuring for Highest Accuracy Measurements
The measurement configurations shown below assume that the multimeter
is in its power
on or reset state. It is also assumed that auto–ranging is
enabled to ensure proper full scale range selection.
DC Voltage, DC Current, and Resistance Measurements:
Select NPLC and 100 (NPLCs) for INTEGRATION.
Set INPUT Z to HI
Z (for the 100 mV, 1 V, and 10 V ranges) for the best dc
voltage accuracy.
Use the 4
wire ohms function (W 4W) for the best resistance
measurement accuracy.
For 2
wire ohms, dc voltage and dc current measurements, set AUTOZERO
to ON to remove thermal EMF and offset errors.
Null the test lead resistance for 2
wire ohms measurements, and to
remove any interconnection offset for dc voltage measurements.
AC Voltage and AC Current Measurements:
Set the AC FILTER to 3 Hz: SLOW.
Frequency and Period Measurements:
Set the GATE TIME to 1 sec.

Table of Contents

Other manuals for Agilent Technologies 34410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34410A and is the answer not in the manual?

Agilent Technologies 34410A Specifications

General IconGeneral
BrandAgilent Technologies
Model34410A
CategoryMultimeter
LanguageEnglish

Related product manuals