34410A/11A/L4411A Service Guide 53
Quick Start 2
Calibration Operation (34410A/11A)
From the front panel you can:
Read the calibration count
Read and set the calibration message.
Secure and unsecure the instrument for calibration.
To Read the Calibration Count
You can query the instrument to determine how many calibrations have
been performed. Note that your instrument was calibrated before it left
the factory. When you receive your instrument, read the count to
determine its initial value. The count value increments by one for each
calibration point, and a complete calibration will increase the value by
many counts.
1 Press .
2 Select CALIBRATION from UTILITY MENU.
3 Select COUNT from CALIBRATION.
4 Make note of the CAL COUNT.