34410A/11A/L4411A Service Guide 19
Specifications 1
DC Characteristics
Accuracy Specifications ±( % of reading + % of range )
[1]
[ 1 ] Specifications are for 90 minute warm–up and integration setting of 100 NPLC.
For <100 NPLC, add the appropriate “RMS Noise Adder” from the table on the following page.
[ 2 ] Relative to calibration standards.
[ 3 ] 20% overrange on all ranges, except 1000 VDC, 3 A range.
[ 4 ] Specifications are for 4–wire ohms function, or 2–wire ohms using Math Null. Without Math Null,
add 0.2 Ω additional error in 2–wire ohms function.
[ 5 ] For each additional volt over ± 500 VDC add 0.02 mV of error.
[ 6 ] Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current
is typical. Variation in the current source will create some variation in the voltage drop across a
diode junction.
Function Range
[3]
Test Current or
Burden Voltage
24 Hour
[2]
T
CAL
± 1 °C
90 Day
T
CAL
± 5 °C
1 Year
T
CAL
± 5 °C
Temperature
Coefficient/°C
0 °C
to (T
CAL
– 5 °C)
(T
CAL
+ 5 °C) to 55 °C
DC 100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V
[5]
0.0030+0.0030
0.0020+0.0006
0.0015+0.0004
0.0020+0.0006
0.0020+0.0006
0.0040+0.0035
0.0030+0.0007
0.0020+0.0005
0.0035+0.0006
0.0035+0.0006
0.0050+0.0035
0.0035+0.0007
0.0030+0.0005
0.0040+0.0006
0.0040+0.0006
0.0005+0.0005
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
0.0005+0.0001
Resistance
[4]
100.0000 Ω
1.000000 KΩ
10.00000 KΩ
100.0000 KΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
1.000000 GΩ
1 mA Current Source
1 mA
100 µA
10 µA
5.0 µA
500 nA
500 nA ||10 MΩ
500 nA || 10 MΩ
0.0030+0.0030
0.0020+0.0005
0.0020+0.0005
0.0020+0.0005
0.0020+0.0010
0.0100+0.0010
0.200+0.001
2.000+0.001
0.008+0.004
0.007+0.001
0.007+0.001
0.007+0.001
0.010+0.001
0.030+0.001
0.600+0.001
6.000+0.001
0.010+0.004
0.010+0.001
0.010+0.001
0.010+0.001
0.012+0.001
0.040+0.001
0.800+0.001
8.000+0.001
0.0006+0.0005
0.0006+0.0001
0.0006+0.0001
0.0006+0.0001
0.0010+0.0002
0.0030+0.0004
0.1000+0.0001
1.0000+0.0001
DC Current 100.0000 µA
1.000000 mA
10.00000 mA
100.0000 mA
1.000000 A
3.00000 A
<0.03 V Burden V
<0.3 V
<0.03 V
<0.3 V
<0.80 V
<2.0 V
0.010+0.020
0.007+0.006
0.007+0.020
0.010+0.004
0.050+0.006
0.100+0.020
0.040+0.025
0.030+0.006
0.030+0.020
0.030+0.005
0.080+0.010
0.120+0.020
0.050+0.025
0.050+0.006
0.050+0.020
0.050+0.005
0.100+0.010
0.150+0.020
0.0020+0.0030
0.0020+0.0005
0.0020+0.0020
0.0020+0.0005
0.0050+0.0010
0.0050+0.0020
Continunity 1000 Ohms 1 mA Test Current 0.002+0.010 0.008+0.020 0.010+0.020 0.0010+0.0020
Diode Test 1.0000 V
[6]
1 mA Test Current 0.002+0.010 0.008+0.020 0.010+0.020 0.0010+0.0020

Table of Contents

Other manuals for Agilent Technologies 34410A