82 34410A/11A/L4411A Service Guide
3 Calibration Procedures
Gain Adjustments
The instrument calculates and stores gain corrections for each input value.
The gain constant is computed from the calibration value entered for the
calibration command and from measurements made automatically during
the adjustment procedure.
Most measuring functions and ranges have gain adjustment procedures.
The 100 MΩ and 1 GΩ ranges do not have gain calibration procedures.
Adjustments for each function should be performed ONLY in the order
shown.
Gain Adjustment Considerations
The ADC and zero adjustment procedure must have been recently
performed prior to beginning any gain adjustment procedures.
Gain adjustments for the 34410A/11A can be made using either the
front or rear input terminals. Make sure the Front/Rear input selection
switch matches the terminals in use.
NOTE
Never turn off the instrument during a Gain Adjustment. This may cause calibration
memory for the present function to be lost.

Table of Contents

Other manuals for Agilent Technologies 34410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34410A and is the answer not in the manual?

Agilent Technologies 34410A Specifications

General IconGeneral
BrandAgilent Technologies
Model34410A
CategoryMultimeter
LanguageEnglish

Related product manuals