34410A/11A/L4411A Service Guide 81
Calibration Procedures 3
Adjust the ADC
3 34410A/11A: Select the front panel input terminals. If using a single
shorting block, install the block on the front panel input terminals.
L4411A: Install the shorting block on the input terminals.
4 Select the ADC adjustment mode (see “Selecting the Adjustment Mode” on
page 78 or, for the remote interface page 79).
5 The display will show the ADC calibration steps as they progress. The
ADC adjustment requires approximately 2 minutes to complete. When
finished, the display will show CAL SUCCEEDED.
Zero Adjustment
6 Select the DMM adjustment mode (see “Selecting the Adjustment Mode” on
page 78). Enter a calibration value of +0.000 E+0 (see “Entering Adjustment
Values” on page 78).
7 The display will show the functions and offset as the adjustments
progress. The Zero Adjustment requires approximately 1.5 minutes to
complete. When finished, the display will show CAL SUCCEEDED.
34410A/11A: Perform the Rear Terminals Zero Adjustment at step 8.
L4411A: skip to step 10.
Rear Terminals Zero Adjustment (34410A/11A Only)
8 Select the rear input terminals. If using a single shorting block, install
the block on the rear panel input terminals.
9 Repeat steps 6 and 7 for the rear input terminals.
Store The Calibration Constants
10 Store the new ADC and Zero adjustment constants (see “Storing the
Calibration Constants” on page 78).
11 Perform the “Zero Offset Verification” on page 64 to check zero calibration
results.