14 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
1 Safety Information
General
WARNING
Warning – High Radio Frequency Radiation
The coil around the torch emits high frequency radiation. Even at low power it can cause
damage to living tissue, therefore the torch must be shielded in a grounded cage. Never
disable the safety features which prevent leakage of High Radio Frequency Radiation via
an open door. High Radio Frequency Radiation can cause interference to medical
electronics such as pacemakers.
Touching the coil during ignition or operation will cause burns.
WARNING
Warning – Exhaust System
Because there can be health hazards, the exhaust gases from the plasma must be
eliminated through the laboratory's ventilation duct. If inadequate ventilation occurs,
ozone, and other toxic products of combustion can accumulate in the laboratory and cause
bodily harm.
Hydrofluoric acid (HF) fumes, if inhaled, can extensively damage lung tissue. Ensure that
the exhaust system established at installation continues to operate effectively. You can
connect a warning light to the extraction system or use an air-flow measurement device to
alert the operator to a malfunction. Prior to operating the instrument, check the
laboratory's ventilation duct.
If the extraction system has an inoperable exhaust fan or insufficient draw, do not start the
plasma. Contact site maintenance personnel.

Table of Contents

Other manuals for Agilent Technologies 5100 ICP-OES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 ICP-OES and is the answer not in the manual?

Agilent Technologies 5100 ICP-OES Specifications

General IconGeneral
BrandAgilent Technologies
Model5100 ICP-OES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals