EasyManuals Logo

Agilent Technologies 5100 ICP-OES Service Manual

Agilent Technologies 5100 ICP-OES
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #298 background imageLoading...
Page #298 background image
298 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
11 Preventative Maintenance
Preventative Maintenance requirements
Preventative Maintenance requirements
Resources required
Personnel required: 1 Trained service engineer
Time to complete: 4 hours
Customer supplied parts:
10 litres clean, distilled water
Multi-element test solution and Blank solution
Anti-algaecide solution (as recommended by the water cooler manufacturer)
Consumable items* Part Number
Nebulizer capillary tubing, per m 2410020500
Peristaltic Pump Tubing white/white (12 pack) 3710034400
Peristaltic Pump Tubing blue/blue (12 pack) 3710034600
TUBE NALGENE 0.12 IDX0.062INTK 3710024600
Tubing s/chamber-torch 1/4 in id, per m 3710033400
Tubing 1/4in OD x 1/8inID PVDF G8000-68003
Tubing-flexible 3mm-ID 4mm-OD 0.5mm-THK 0890-2603
FERRULE- 6MM OD TUBE- SWAGELOK 1610119800
Concentric Glass Nebulizer Seaspray UF G8010-60255
Glass s/chamber d/p helix ball joint UFT G8010-60256
Hose for cooling water, 10m G1833-65429
ICP-OES Easy Fit DV Torch G8010-60228
ICP-OES Easy Fit Radial Torch G8012-60000
Filter Argon ICP-OES 5110 G8010-60136

Table of Contents

Other manuals for Agilent Technologies 5100 ICP-OES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 ICP-OES and is the answer not in the manual?

Agilent Technologies 5100 ICP-OES Specifications

General IconGeneral
BrandAgilent Technologies
Model5100 ICP-OES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals