96 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Igniter
4 Loosen coil nuts and remove coil.
5 Remove nut and insulator that secures current sensor to SSRF body. (Do not overtighten this
nut when reassembling!)
6 Slide current sensor away from coil shaft and replace as required.
NOTE
The sensor fits tightly on a morse taper and may need a gentle tap to remove.
Igniter
Igniter replacement requires removal of the RF Oscillator as described in the previous section.