24 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
3 General Information
Specifications
Specifications
Environmental Conditions and Utility Requirements
Return to step 8
Agilent 5110 ICP-OES Specification
Power requirements
Power (kVA) 2.9 kVA
Voltage (AC) 200-240V single phase
Mains Frequency 50-60Hz
Mains Wall socket required (Australia) 15Amp GPO
Gas requirements
Argon max consumption (L/min) 35.5 L/min
Argon working pressure (kPa) 500-600kPa (73-88 psi) gauge
Argon recommended operating pressure 550kPa (80 psi)
Argon instrument hose fitting quick connect
Argon gas regulator fittings quick connect
Argon purity 99.995%
Argon oxygen content (ppm) <5ppm
Argon Nitrogen content (ppm) <20ppm
Argon water vapor (ppm) <4ppm
Nitrogen max consumption (optics purge) 3.7 L/min
Nitrogen working pressure (kPa) 500-600 kPa (73-88 psi)
Nitrogen recommended operating pressure 550 kPa (80 psi)
Nitrogen instrument hose fittings quick connect
Nitrogen gas regulator fittings quick connect
Nitrogen Purity 99.995%
Nitrogen oxygen content <5ppm
Nitrogen water vapor (ppm) <4ppm
Thermal Extraction
Exhaust Air Extraction 2.5 m
3
/min > flow < 6.0 m
3
/min minimum
(90 ft
3
/min > flow < 200 ft
3
/min)
Water cooling inlet recommended operating temperature (°C) 20°C
Water cooling inlet temperature max (°C) 28°C
Water flow rate (minimum) (L/min) 1.7 L/min
Water minimum cooling inlet temperature 15°C
Water cooling minimum inlet pressure (kPa) 230kPa (33 psi)