EasyManuals Logo

Agilent Technologies 5100 ICP-OES Service Manual

Agilent Technologies 5100 ICP-OES
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #153 background imageLoading...
Page #153 background image
Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 153
4 Removal/ Installation, Replacement and Adjustment
Argon ratio adjustment
13 Adjust DLA screw in the opposite direction from step 12 until the argon ratio equals 4.5 to
5.0. (This determines the maximum values for each of the 12 selected elements). Adjust the
screw in 0.2 increments allowing the normalized intensity graphs to update until a flat
response is met. (All 12 elements are displayed at approximately 90% intensity or better).
Record the argon ratio on the Status page, (more instrument settings section) when the
registration graph shows the best response for all 12 elements.
14 Replace all polychromator access plugs and covers.
15 Go to the Expert 7 software and perform a dark current scan.
16 Perform a wavelength calibration.

Table of Contents

Other manuals for Agilent Technologies 5100 ICP-OES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 ICP-OES and is the answer not in the manual?

Agilent Technologies 5100 ICP-OES Specifications

General IconGeneral
BrandAgilent Technologies
Model5100 ICP-OES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals