72 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Top right hand cover
4 Remove torx 10 screw from front right of top cover.
5 Remove torx 10 screw from front left of top cover.
6 Remove torx 10 screw from rear left of top cover.
7 Lift top cover, push to rear of instrument to separate rear locking tabs and then remove by
sliding cover towards the right of the instrument.