Maintenance
100 Agilent 55B AA Spectrometer User’s Guide
Do not force the white plastic insert. It may be necessary to screw
counter-clockwise first for lead-in thread to be located.
8 Place the clamp bush guide, thread end first, in the nebulizer body.
Figure 50. Insert the clamp bush guide into the Nebulizer block.
9 Screw the clamp bush guide into the white plastic insert using the small end of
the Nebulizer tool. Do not over-tighten.
10 Check the spring is operational by pushing the capillary assembly with a screw
driver. The assembly should move in and out freely.
Figure 51. Check spring function
11 Thread the thimble and locking ring over the capillary.
12 Screw the thimble onto the nebulizer body.
Instructions on how to adjust the Nebulizer are detailed in ‘Nebulizer setup’ on
Page 81.