Maintenance
Agilent 55B AA Spectrometer User’s Guide 121
Figure 83. Polish the
outer surface of the burner slot
15 Rinse the burner with distilled water.
16 Wash and rinse the burner with a diluted 10% v/v solution of an alkali
detergent to remove any traces of the metal polishing liquid.
17 Dry the burner. Use compressed air to fasten the drying process.
NOTE
The top of the burner slot should be clean and shiny to indicate that all deposits
have been cleaned off.
18 If necessary, repeat some or all the previous steps until the burner slot is
clean.
Table 9. Maximum allowable burner slot width
N
2
O
Air
Mark VIA
0.47 mm (0.0185 in.)
Mark 7
0.46 mm (0.0181 in.)
0.54 mm (0.021 in.)
Burners cannot be refurbished and must be replaced when the maximum slot
widths are exceeded.
NOTE
For applications with high deposits, additional burner cleaning procedures are
recommended. Contact your Agilent customer support representative for
recommendations.