4 Operating in EI Mode
78 5975/5977 Series MSD Operating Manual
To change MSD temperatures from the GC
control panel
1 Press MS/Aux Det to display the 5977B MSD menu.
2 Press the down arrow to scroll to Quad temp, Source temp or Transfer line.
3 Use GC keypad to enter the desired temperature.
4 Press Enter to apply the changes.
To view MSD vacuum pressure and Turbo
speed/Foreline Pressure from the GC control
panel
1 Press MS/Aux Det to display the 5977B MSD menu.
2 Press the down arrow to scroll to HiVac Pressure, or Turbo Speed % of full
/ Foreline Pressure.
To vent the MSD from the GC control panel
1 With the MSD pumped down, press MS/Aux Det to bring up the 5977B MSD
menu.
2 Press the down arrow to scroll to Start MSD Vent?. (Press Off/No to cancel
the vent cycle and pump down the MS).
3 Press ON/Yes to start the venting cycle.
4 When prompted open the vent valve.
To pump down the MSD from the GC control
panel
1 With the MSD vented, press MS/Aux Det to bring up the 5977B MSD menu.
2 Press the down arrow to scroll to Start MSD Vent?.
3 Press ON/Yes to start the pump down cycle.
4 When prompted open the vent valve.

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 Series and is the answer not in the manual?

Agilent Technologies 5975 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals