4 Operating in EI Mode
5975/5977 Series MSD Operating Manual 79
To view the firmware version of the MSD from
the GC control panel
1 Press MS/Aux Det to bring up the 5977B MSD menu.
2 Press the down arrow to scroll to Firmware.
To view the serial number of the MSD from the
GC control panel
1 Press MS/Aux Det to bring up the 5977B MSD menu.
2 Press the down arrow to scroll to Serial#.
To configure the network settings for the MSD
from the GC control panel
1 Press Config, and then press MS/Aux Det to bring up the CONFIGURE MS
DETECTOR menu.
2 To configure the IP: parameter, use the GC keypad to enter the new IP
address for the MSD, then press Enter to complete the entry.
3 Wait for the GC to display the new IP address. Reboot the MSD or proceed to
the gateway address with the down arrow button
4 Press the down arrow to scroll to GW: and use the GC keypad to enter the
new gateway address for the LAN and press Enter to complete the entry.
5 Press the down arrow to scroll to SW: and use the GC keypad to enter the
new subnet mask for the LAN and press Enter to complete the entry.
6 Reboot the MSD. (See below)
To reboot the MSD from the GC control panel
1 Press Config, and then press MS/Aux Det to bring up the CONFIGURE MS
DETECTOR menu.
2 Press the down arrow to scroll to Request MSD Reboot?.
3 Press On/Yes to reboot the MSD and wait for the MSD to complete this cycle
before trying to access it.