4 Basic Operation
Setting Up and Running an Acquisition Method
Agilent 6475 Triple Quad LC/MS System User Guide 54
6 Set up the Stored instrument curves in the Instrument tab. In the Qualitative
Analysis program, you can display these values in the MS Actuals window for
the current spectrum. With the instrument, the values in the MS Actuals
window in the Qualitative Analysis program are the values that you save in the
Instrument tab.
a Click Instrument.
b Select the Stored instrument curves. These curves can be shown in the
Chromatogram Results window in the Qualitative Analysis program. The
values can be seen in the MS Actuals window.
7 Set up the data analysis (DA) parameters.
A method can contain Qualitative Analysis parameters, Quantitative Analysis
parameters, or a combination. The Qual Automation check box is marked on
the Qual tab and the Quant Automation check box is marked on the Quant
tab.
a Select the DA tab.
b Optional. On the Qual tab, mark the Qual Automation check box. The name
of the current Qualitative Analysis method is shown in the box. If you want
to change the Qualitative Analysis method that is connected, click browse
to select a different method. When the Data Acquisition method is saved,
the Qualitative Analysis method that you selected is copied or linked to the
Data Acquisition method.