5 Maintenance
Sources
Agilent 6475 Triple Quad LC/MS System User Guide 68
Removing an APCI source
1 Put the system in Standby mode.
NOTE
In the Tune view or context, turn down the gas temperature as low as possible.
Leave the gas flow on for 30 minutes to cool the source.
2 Disconnect the nebulizer sample line.
3 Disconnect the sample nebulizer gas tubing.
4 Disconnect the heater cable.
5 Disconnect the corona needle voltage cable.
Removing a Mutlimode source
1 Put the system in Standby mode.
NOTE
In the Tune view or context, turn down the gas temperature as low as possible.
Leave the gas flow on for 30 minutes to cool the source.
2 Disconnect the nebulizer sample line.
3 Disconnect the heater cable.
4 Disconnect the charging electrode cable.
5 Open the latch of spray chamber.
WARNING
The spray chamber operates at very high temperature. Do not continue until
the spray chamber is cool.
6 Remove the spray chamber by lifting it along the axis of its hinges.
WARNING
The spray shield can be hot. Be careful not to burn yourself when you remove
the spray shield.
7 If you are changing to a different source type, remove the spray shield.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals