5 Maintenance
Sources
Agilent 6475 Triple Quad LC/MS System User Guide 71
When all set points are equilibrated:
1 In the Tune view or context, turn on the calibrant.
2 Purge the calibrant line for 5 minutes.
3 Do a calibration. If calibration is not available, do a check tune.
4 If the check tune results are not acceptable, run auto tune on a supported
tuning source.
5 When the instrument is ready, load or create the method to use with the
source.
6 Change the method settings for the new source.
Installing a Mutlimode source
1 Install the appropriate spray shield.
2 Make sure the dowels are at 6 and 9 o'clock positions.
3 Install the spray chamber.
4 Close the spray chamber.
5 Close the latch of spray chamber.
6 Connect the corona needle voltage cable.
7 Connect the charging electrode cable.
8 Connect the heater cable.
9 Connect the sample nebulizer gas tubing.
10 Connect the nebulizer sample line.
In the Data Analysis program,
1 In the Tune context, select the new source.
2 Turn on the LC/MS instrument.
3 While you wait for the set points to equilibrate, prepare and install calibrant for
the new source.
a Rinse a clean calibrant bottle with LC/MS-grade acetonitrile.
b Pour the MMI Calibrant into the calibrant bottle.
c Install the calibrant bottle into the calibrant delivery system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals