EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6475

Agilent Technologies 6475 User Manual

Agilent Technologies 6475
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
5 Maintenance
Nebulizer
Agilent 6475 Triple Quad LC/MS System User Guide 77
Installing the nebulizer
1 Insert the nebulizer part way into the spray chamber.
CAUTION
Do not hit the tip of the needle as you insert the nebulizer. The tip of the needle
is easily damaged.
2 Connect the nebulizer sample line to the nebulizer.
3 Turn the nebulizer clockwise and lock it in place.
4 Connect the sample nebulizer gas tubing to the nebulizer.
CAUTION
Do not over-tighten the LC fitting. An over-tightened fitting can crush the
tubing, which creates a restriction.
Fixing Spray Stability EMF
If spray stability EMF is active, you should inspect, adjust or flush your nebulizer
to ensure a stable spray from the CDS to the nebulizer. An unstable spray may
lead to issues with tuning and sample analysis.
1 Inspect (page 73), adjust (page 75), or flush page 65) your nebulizer. If
necessary, replace the nebulizer needle or nebulizer (page 75).
2 Follow the procedure to “Calibrating the LC/TQ (Checktune)” on page 30 or
“Calibrating the LC/TQ (Autotune)” on page 31. The spray stability is
reevaluated as part of a tune. If it within the acceptable range, the EMF
warning should be removed following completion of the tune.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals