Chapter 9 693
If You Have a Problem
Error Messages
Error codes 300 through 399 relate to LO and RF
hardware/firmware failures. Instrument service is required.
ERR 300  YTO UNLK YTO (1ST LO) phase-locked loop (PLL) is
unlocked.
ERR 301  YTO UNLK YTO PLL is unlocked.
ERR 302  OFF UNLK Offset roller oscillator PLL is unlocked.
ERR 303  XFR UNLK Transfer roller oscillator PLL is unlocked.
ERR 304  ROL UNLK Main roller oscillator PLL is unlocked.
ERR 305  FREQ ACC Coarse adjust DAC cannot bring MAINSENSE
close to zero.
ERR 306  FREQ ACC Fine adjust DAC cannot bring MAINSENSE
close to zero.
ERR 307  FREQ ACC Transfer oscillator pretune DAC out of range.
ERR 308  FREQ ACC Offset oscillator pretune DAC not within
prescribed limits at low frequency.
ERR 309  FREQ ACC Offset oscillator pretune DAC not within
prescribed limits at high frequency.
ERR 310  FREQ ACC Main roller pretune DAC set to 255 before
MAINSENSE changes to negative polarity.
ERR 311  FREQ ACC Main roller pretune DAC set to 255 before
MAINSENSE changes to negative polarity.
ERR 312  FREQ ACC Fine adjust DAC cannot bring MAINSENSE
close to zero.
ERR 313  FREQ ACC The combination of sampler oscillator and
roller oscillator frequencies do not correspond to the
required YTO start frequency.
ERR 314  FREQ ACC Span calibration problems. An unlocked main
roller loop or lack of a sweep ramp.
ERR 315  FREQ ACC Span calibration problems. Roller span
attenuator DAC is out of range.
ERR 316  FREQ ACC Sensitivity of the main roller DAC is zero.
ERR 317  FREQ ACC Main coil coarse DAC at limits.
ERR 318  FREQ ACC Main coil fine DAC at limits.
ERR 319  WARN COA The YTO coarse tune DAC is near its limits.
ERR 320  WARN FIN The YTO fine tune DAC is near its limits.