694 Chapter9
If You Have a Problem
Error Messages
ERR 321  FREQ ACC Main roller tuning sensitivity is not greater
than zero.
ERR 322  FREQ ACC Main roller pretune DAC value set greater
than 255.
ERR 324  FREQ ACC Coarse adjust DAC cannot bring MAINSENSE
close to zero.
ERR 325  FREQ ACC Fine adjust DAC cannot bring MAINSENSE
close to zero.
ERR 326  FREQ ACC Fine adjust DAC near the end of range.
ERR 327  OFF UNLK Offset roller oscillator PLL is unlocked.
ERR 328  FREQ ACC Roller fine adjust DAC sensitivity less than or
equal to zero.
ERR 329  FREQ ACC Roller coarse adjust DAC sensitivity less than
or equal to zero.
ERR 331  FREQ ACC Invalid YTO frequency.
ERR 333  600 UNLK 600 MHz reference oscillator PLL is unlocked.
ERR 334  LO AMPL 1st LO distribution amplifier is unleveled. The
front panel LO OUTPUT must be terminated in a 50Ω
load.
ERR 335  SMP UNLK Sampling oscillator PLL is unlocked.
ERR 336  10 MHz Ref Cal oscillator failed to lock when going to
internal 10 MHz reference.
ERR 337  FN UNLK Fractional N circuitry is unable to lock.
ERR 338  LOMA AGC The LOMA (local oscillator multiplier
amplifier) main loop AGC is unleveled.
ERR 339  SBTX AGC The LOMA SBTX (switched barium-tuned
mixer) loop AGC is unleveled.
ERR 340  SAMP AGC The LOMA (local oscillator multiplier
amplifier) sampler loop AGC is unleveled.
ERR 351  SETL FLD The YTO error voltage is not settling.
ERR 352  TWID FLD Unable to bring error voltage DACs to
quiescent point.
ERR 353  SRCH FLD No acceptable YTO DAC value found.
ERR 354  LK ITERS Cannot lock. Lock iteration routine
terminated.
ERR 355  SMP CAL Sampler unlock condition during calibration
routine. This error remains until a successful
recalibration is performed.