11-16 Programmer’s Guide
SCPI Command Summary
SCPI Device Command Summary
Table 11-6 DIAGnostic (3 of 4)
SUBSYSTEM COMMANDS FORM DESCRIPTION
DIAGnostic:MDISplay[1|2]
:CORRection <string>
command
only
Displays corrected measurement
uncertainties. Choose from one of the
following strings:
Cal check
C_DIRECTivity
C_LDMATCH
C_ISOLATION
C_RTRACKING
C_SRCMATCH
C_TTRACKING
Interpolated Array (accessed through the
service menu.)
I_DIRECTivity
I_RESPONSE
I_SRCMATCH
I_TRACKING
I_LDMATCH
I_ISOLATION
Master Array (accessed through the
service menu.)
M_DIRECTivity
M_RESPONSE
M_SRCMATCH
M_TRACKING
M_XSCALAR
M_LDMATCH
I_ISOLATION
DIAGnostic:MDISplay[1|2]
:RESTore
command
only
Returns to measurement mode and
autoscales after viewing calibration
uncertainties.