EasyManua.ls Logo

Agilent Technologies 8712ET - Page 261

Agilent Technologies 8712ET
320 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programmer’s Guide 11-63
SCPI Command Summary
SCPI Device Command Summary
Table 11-23 SYSTem (5 of 8)
1. For more information on errors, refer to Chapter 13, “SCPI Error Messages.
2. A list analyzer front panel key codes is provided in Chapter 8, “Front Panel Key-
codes.
SUBSYSTEM COMMANDS FORM
DESCRIPTION
SYSTem:COMMunicate:SERial
:TRANsmit:BAUD <num>
NR1 Sets the baud rate for hardcopy
output to a device on the serial port
— choose from
1200|2400|4800|9600|19200.
SYSTem:COMMunicate:SERial
:TRANsmit:HANDshake <char>
CHAR Sets the handshake for
communication to a hardcopy device
on the serial port — choose XON or
DTR.
SYSTem:COMMunicate:TTL:USER:
FEED <char>
CHAR Selects the function of the USER
TTL IN/OUT port on the rear panel
of the analyzer. Choose from
DEFault|KEY|SWEep.
SYSTem:DATE
<num1>,<num2>,<num3>
NR1,
NR1,
NR1
Sets the year (num1), month (num2)
and day (num3) of the real time
clock.
SYSTem:ERRor?
1
query
only
NR1,
STRING
Queries the error queue — returns
the error number and message.
SYSTem:KEY <char> command
only
Sends key names
2
which execute
the same functions as front panel
keys.