4-45
Screens
Cell Control 2
31. SACCH Meas These fields indicate the measurement results of a slow associated control channel
measurement.
TX Lev indicates the mobile station’s reported transmitter power level.
Tim Adv is the mobile station’s reported timing advance.
Full RX Lev is the mobile station’s received level of power from the serving cell
using the Full measurement method.
Full RX Qual is the mobile station’s received quality of signal from the serving
cell using the Full measurement method.
Partial RX Lev is the mobile station’s level of power from the serving cell using
the Partial measurement method.
Partial RX Qual is the mobile station’s received quality of signal from the
serving cell using the Partial measurement method.
Adj Cell RX Lev is the reported level of power of the adjacent cell selected in the
Adj Cell field.
Adj Cell ARFCN is the absolute radio frequency channel number of the adjacent
cell selected in the Adj Cell field.
BSIC NCC reads the NCC: PLMN Colour-Code portion of the base station
identity code (BSIC) of the adjacent cell selected in the Adj Cell field.
BSIC BCC reads the BCC: Base Station Colour-Code portion of the base station
identity code (BSIC) of the adjacent cell selected in the Adj Cell field.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8922M and is the answer not in the manual?

Agilent Technologies 8922M Specifications

General IconGeneral
BrandAgilent Technologies
Model8922M
CategoryTest Equipment
LanguageEnglish

Related product manuals