EasyManua.ls Logo

Agilent Technologies 8922M - Page 311

Agilent Technologies 8922M
381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-15
Instrument BASIC
Loading, Storing, and Running
Figure 8-5 Making a Test Procedure File
Running a Test
Sequence
1 Select the Run Test field.
2 Follow directions and prompts on the Agilent 8922M/S screen according to the
test sequence being run.
3 When testing is complete, the Agilent 8922M/S will respond to front panel or
remote input. If at any time you need to stop testing, press the front-panel
key.
NOTE A program takes up to 3 minutes to load into the Agilent 8922M/S and is loaded when
the Run Test field is first pressed.
Press the front-panel keys to abort from an error condition or to
abort from the program. When you abort from loading the program, you’ll need to
clear Agilent 8922M/S RAM memory in order to correctly re-load the program at a
later time.
To clear Agilent 8922M/S RAM, select and run the ROM program COPY_PL from
the Procedure field in the TESTS menu. (Running COPY_PL deletes all
SAVE/RECALL registers. So instead you may want to load another program, run it,
and then re-load the original program that was aborted.
CANCEL
SHIFT CANCEL

Table of Contents

Related product manuals