2-8
Making Measurements
ACTIVE CELL
There may be many reasons for the measured level not being close to the expected
level. The two most likely are that, either, the mobile phone is not operating
correctly, or, there is some power loss between the Agilent 8922M/S RF IN/OUT
connector and the mobile phone. If you suspect it is the second case, you can
compensate the Agilent 8922M/S generator settings and measurement results for
external losses or gains. The compensation is carried out in the CONFIGURE screen
which is accessed using the CONFIG field on the bottom right-hand side of the
screen, (6), see Figure 2-2 on page 2-6. Refer to Chapter 4, Screens for further
information.
Timeslot To change the timeslot, highlight the field and enter a new timeslot value
from the keypad. The timeslots can vary from 2 to 6. (Timeslots 0, 1, and 7 are
reserved for maintaining communication between the Agilent 8922M/S and the
mobile phone.)
NOTE While the Active Cell is selected, the three MOBILE PHONE parameters are also
available on the bottom right-hand side of all measurement screens. This provides
control of the mobile phone during measurements.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8922M and is the answer not in the manual?

Agilent Technologies 8922M Specifications

General IconGeneral
BrandAgilent Technologies
Model8922M
CategoryTest Equipment
LanguageEnglish

Related product manuals