3-18
Verifying Performance
Agilent Technologies 8922M/S Specifications
RF Analyzer
Specifications
Frequency
CW RF Frequency Measurement
CW RF Power Measurements (RF In/Out Only)
Range 10 MHz to 1 GHz
Resolution 1 Hz
Hop Mode
Resolution
100 kHz
Offset Frequency
50 kHz
Offset Resolution
1 Hz
RF In/Out SWR <1.5:1
Supplemental Characteristics:
Frequency Overrange
To 1015 MHz
Offset Resolution
500 Hz for FM demodulation out
Range 10 MHz to 1 GHz
Level Range
RF In/Out
6 to +41 dBm
Aux RF In
36 to +20 dBm
Input Frequency Setting Error
±500 kHz
Accuracy
±(1 Hz + reference accuracy)
Supplemental Characteristics:
Minimum Resolution
1 Hz
Range -5 to +41 dBm
Input Frequency Setting Error
±500 kHz
Accuracy
±0.5 dB ± noise effects (+4 to +41 dBm) (0.2 mW)
Supplemental Characteristics:
Accuracy
±0.5 dB ± noise effects (-5 to +4 dBm) (0.2 mW)
Minimum Resolution
0.01 dB