LAN Configuration
l This setting is non-volatile; it will not be changed by power cycling or *RST.
l This is set to "33521A> " (Agilent 33521A), "33522A> " (Agilent 33522A), or "33500> " (other 33500 Series
instruments) when the instrument is shipped from the factory or after SYSTem:SECurity:IMMediate.
SYSTem:COMMunicate:LAN:TELNet:WMESsage "<string>"
SYSTem:COMMunicate:LAN:TELNet:WMESsage?
Sets welcome message seen when communicating with instrument via Telnet.
Parameter Typical Return
String of up to 63 characters "Welcome to the Telnet Session"
Define a welcome message:
SYST:COMM:LAN:TELN:WMES "Welcome to the Telnet Session"
l This setting is non-volatile; it will not be changed by power cycling or *RST.
l This is set to "Welcome to Agilent's 33521A Waveform Generator" (Agilent 33521A), "Welcome to Agilent's
33522A Waveform Generator" (Agilent 33522A), or "Welcome to Agilent's 33500-Series Waveform Generator"
(other 33500 Series instruments) when the instrument is shipped from the factory or after SYS-
Tem:SECurity:IMMediate.
SYSTem:COMMunicate:LAN:UPDate
Stores any changes made to the LAN settings into non-volatile memory and restarts the LAN driver with the updated
settings.
Parameter Typical Return
(none) (none)
(see below)
l This command must be sent after changing the settings for DHCP, DNS, gateway, hostname, IP address, subnet
mask, WINS.
l Make all changes to the LAN settings before sending this command.
Example
The following example configures the instrument to use statically assigned LAN settings.
SYST:COMM:LAN:DHCP OFF
SYST:COMM:LAN:DNS "198.105.232.4"
SYST:COMM:LAN:DNS2 "198.105.232.5"
SYST:COMM:LAN:GATEWAY "198.105.232.1"
SYST:COMM:LAN:HOST "LAB1-33522A"
SYST:COMM:LAN:IPAD "198.105.232.101"
SYST:COMM:LAN:SMAS "255.255.255.0"
SYST:COMM:LAN:WINS "198.105.232.4"
Agilent 33500 Series Operating and Service Guide 313