198 Chapter 5
Language Reference
ABORt Subsystem
ABORt Subsystem
Abort
:ABORt
Restarts any sweep or measurement in progress and resets the sweep or trigger
system. A measurement refers to any of the measurements found in the
MEASURE
menu.
If :INITiate:CONTinuous is off (single measure), then
:INITiate:IMMediate will start a new single measurement.
If :INITiate:CONTinuous is on (continuous measure), a new continuous
measurement begins immediately.
The INITiate and TRIGger subsystems contain additional related commands.
Front Panel
Access:
Restart for continuous measurement mode