EasyManua.ls Logo

Agilent Technologies E7401A

Agilent Technologies E7401A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
316 Chapter 5
Language Reference
STATus:QUEStionable Subsection
Questionable Integrity Uncalibrated Event Query
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?
This query returns the decimal value of the sum of the bits in the Questionable
Integrity Uncalibrated Event register.
NOTE The register requires that the equivalent PTR or NTR filters be set before a
condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the data is
cleared.
Questionable Integrity Uncalibrated Negative Transition
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition
<integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition?
This command determines which bits in the Questionable Integrity Uncalibrated
Condition register will set the corresponding bit in the Questionable Integrity
Uncalibrated Event register when that bit has a negative transition (1 to 0). The
variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Factory Preset
and *RST: 0
Range: integer, 0 to 32767
Questionable Integrity Uncalibrated Positive Transition
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition
<integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition?
This command determines which bits in the Questionable Integrity Uncalibrated
Condition register will set the corresponding bit in the Questionable Integrity
Uncalibrated Event register when that bit has a positive transition (0 to 1). The
variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Factory Preset
and *RST: 32767 (all 1s)
Range: integer, 0 to 32767