EasyManuals Logo

Agilent Technologies E7401A Programmer's Guide

Agilent Technologies E7401A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #291 background imageLoading...
Page #291 background image
Chapter 5 291
Language Reference
[:SENSe]:FREQuency Subsection
Center Frequency Step Size
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Specifies the center frequency step size.
Factory Preset
and *RST: Span/10
Range: Maximum negative frequency to the maximum positive
frequency listed below:
EMC E7401A: 1.58 to 1.58 GHz
EMC E7402A: 3.10 to 3.10 GHz
EMC E7403A: 6.78 to 6.78 GHz
EMC E7404A: 13.3 to 13.3 GHz
EMC E7405A: 27.0 to 27.0 GHz
Default Unit: Hz
Front Panel
Access:
FREQUENCY/Channel, CF Step Man
Frequency Span
[:SENSe]:FREQuency:SPAN <freq>
[:SENSe]:FREQuency:SPAN?
Set the frequency span. Setting the span to 0 Hz puts the analyzer into zero span.
Factory Preset
and *RST: 800 MHz
Range: EMC E7401A: 0 Hz, 100 Hz to 1.58 GHz
EMC E7402A: 0 Hz, 100 Hz to 3.10 GHz
EMC E7403A: 0 Hz, 100 Hz to 6.78 GHz
EMC E7404A: 0 Hz, 100 Hz to 13.3 GHz
EMC E7405A: 0 Hz, 100 Hz to 27.0 GHz
Default Unit: Hz
Front Panel
Access:
SPAN/X Scale, Span
SPAN/X Scale, Zero Span

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals