EasyManua.ls Logo

Agilent Technologies E7401A

Agilent Technologies E7401A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 335
Language Reference
TRIGger Subsystem
Factory Preset
and *RST: Positive
Front Panel
Access:
Trig, External Pos Neg
Trigger Offset
:TRIGger[:SEQuence]:OFFSet <64 bit floating point value>
:TRIGger[:SEQuence]:OFFSet?
This command sets the trigger offset.
Factory Preset
and *RST: 0 seconds
Example: :TRIG:SEQ:OFFS 1.0s
Range: Hardware specific; dependent upon the ADC being used,
current state, and the number of sweep points.
Default Unit: seconds
Remarks: Trigger offset refers to the specified time interval before or after
the trigger event from which data is to be written to the trace,
and then displayed. Ordinarily, the trigger offset value is zero,
and trace data is displayed beginning at the trigger event. A
negative trigger offset value results in the display of trace data
prior to the trigger event. A positive trigger offset value results
in an effective delay in the display of trace data after the trigger
event.
The trigger offset value used when the feature is enable will
depend on the following parameters:
Normal trigger offset value originally entered
Specific instrument hardware in use
Sweep time
Number of sweep points
The effective trigger offset value will be re-calculated whenever
any of these parameters change.
Front Panel
Access:
Trig, Trig Offset

Table of Contents

Other manuals for Agilent Technologies E7401A

Related product manuals