Chapter 5 237
Language Reference
DISPlay Subsystem
Window Annotation
:DISPlay:WINDow:ANNotation[:ALL] OFF|ON|0|1
:DISPlay:WINDow:ANNotation[:ALL]?
Turns the screen annotation on or off for all windows.
Factory Preset
and *RST: On
Front Panel
Access:
Display, Preferences, Annotation On Off
Trace Graticule Display
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow:TRACe:GRATicule:GRID[:STATe]?
Turns the graticule on or off.
Factory Preset
and *RST: On
Front Panel
Access:
Display, Preferences, Graticule On Off
Trace X-Axis Scale Offset
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet <freq>
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet?
Specifies the frequency offset for all frequency readouts such as center frequency,
except that it does not affect marker count.
Factory Preset
and *RST: 0 Hz
Range: 500 THz to 500 THz
Default Unit: Hz
History: Prior to firmware revision A.06.00, the lower range is 3 GHz.
Remarks: Frequency offset is not available when frequency scale type is
Log ([:SENSe]:SWEep:SPACing LINear|LOGarithmic).
Front Panel
Access:
FREQUENCY/Channel, Freq Offset

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals