280 Chapter 5
Language Reference
[:SENSe]:DEMod Subsection
[:SENSe]:DEMod Subsection
Type of Demodulation
[:SENSe]:DEMod AM|FM
[:SENSe]:DEMod?
Sets the type of demodulation.
Factory Preset
and *RST: AM
Front Panel
Access:
Det/Demod, Demod, AM
Det/Demod, Demod, FM
FM Deviation
[:SENSe]:DEMod:FMDeviation <freq>
[:SENSe]:DEMod:FMDeviation?
Sets the total FM frequency deviation for full screen demodulation.
Factory Preset
and *RST: 100 kHz
Range: 5 kHz to 1.2 MHz
Default Unit: Hz
Front Panel
Access:
AMPLITUDE, Scale/Div
Squelch
[:SENSe]:DEMod:SQUelch <integer>
Sets the squelch level on FM demod.
Factory Preset
and *RST: integer, 0 to 100
Key Access:
Det/Demod, Demod/Audio, Squelch