Chapter 5 253
Language Reference
INPut Subsystem
Clear the Input Overload
:INPut:PROTection:CLEar
Resets the overload protection circuitry for the input connector. There is no query
form of this command.
NOTE This command is valid only for Agilent EMC model E7401A.
The excessive input signal may have caused 15 dB of attenuation to be switched
in, or it may have completely switched the input connector out so that it is
connected to the internal reference signal.