EasyManuals Logo

Agilent Technologies E7401A Programmer's Guide

Agilent Technologies E7401A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #272 background imageLoading...
Page #272 background image
272 Chapter 5
Language Reference
[:SENSe]:BANDwidth Subsection
[:SENSe]:BANDwidth Subsection
Resolution Bandwidth
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
Specifies the resolution bandwidth.
Example: BAND 1 kHz
Range: 10 Hz to 5 MHz (with Option 1DR)
1 Hz to 5 MHz (E7401A, E7402A, E7403A, E7404A, E7405A
with Option 1D5)
Default Unit: Hz
Front Panel
Access:
BW/Avg, Resolution BW Auto Man
Resolution Bandwidth Automatic
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
Couples the resolution bandwidth.
Factory Preset
and *RST: On
Example: BWID:AUTO On
History: This command function changed with firmware revision
A.08.00. With :AUTO ON in zero span, an error will be
generated.
Remarks: Auto-couple resolution bandwidth is not available in zero span.
Resolution Bandwidth Mode
[:SENSe]:BANDwidth|BWIDth[:RESolution]:MODE EMI|SAN|OFF
[:SENSe]:BANDwidth|BWIDth[:RESolution]:MODE?
Couples the resolution bandwidth to the frequency span in SAN mode, to Center
Frequency in EMI mode, or uncoupled when OFF.
Factory Preset
and *RST: EMI

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals