Chapter 5 299
Language Reference
[:SENSe]:SWEep Subsection
Factory Preset
and *RST: Linear
Remarks: Refer to the Users Guide for detailed information on the
interactions of this command with other functions.
History: Added with firmware revision A.08.00.
Front Panel
Access:
FREQUENCY, Scale Type
Sweep Time
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the display.
Factory Preset
and *RST: 120.8 ms
Range: The range depends upon the installed options, number of sweep
points, and firmware revision of your instrument. See Sweep
Time Range in the Specifications Guide for details.
Default Unit: seconds
Remarks: A span value of 0 Hz causes the analyzer to enter zero span
mode. In zero span the X-axis represents time rather than
frequency. In this mode, the sweep time may be set to faster
values when Option AYX is installed.
Front Panel
Access: Sweep, Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current settings.
Factory Preset
and *RST: On
History: This command function changed with firmware revision
A.08.00. With :AUTO ON in zero span, an error will be
generated.
Front Panel
Access:
Sweep, Sweep Time Auto Man

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals