EasyManuals Logo

Agilent Technologies E7401A Programmer's Guide

Agilent Technologies E7401A
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #269 background imageLoading...
Page #269 background image
Chapter 5 269
Language Reference
[:SENSe]:AVERage Subsection
[:SENSe]:AVERage Subsection
Clear the Current Average
[:SENSe]:AVERage:CLEar
Re-start the trace averaging function.
NOTE Re-start the trace at the beginning of a sweep to obtain valid average data. To
do this, remotely abort the sweep and initiate a single sweep.
Set the Average Count
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Specifies the number of measurements that are combined.
Factory Preset
and *RST: 100
Range: 1 to 8192
Front Panel
Access:
BW/Avg, Average On Off
Turn Averaging On/Off
[:SENSe]:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:AVERage[:STATe]?
This command toggles averaging off and on. Averaging combines the value of
successive measurements to average out measurement variations.
Factory Preset
and *RST: Off
Remarks: When a measurement under the front panel
MEASURE key is
started, this command is turned off for video averaging
([:SENSe]:AVERage:TYPE VIDeo). If this command is
turned on for video averaging when any of the MEASURE key
measurements are in progress, that measurement will be
stopped.
Front Panel
Access:
BW/Avg, Average On Off

Table of Contents

Other manuals for Agilent Technologies E7401A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E7401A and is the answer not in the manual?

Agilent Technologies E7401A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE7401A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals